Comark C9501/IS Pressure Meter 0 to ± 140mbar, 2 PSI

€730.00
-
+
€730.00

HH Solutions supply the Comark Pressure Meter Range in Ireland including the C9501/IS ATEX Compliant Intrinsically Safe Model.

The Comark C9501/IS Intrinsically Safe Pressure Meter with a range of 0 to ± 140mbar / 2 PSI, it is an Intrinsically Safe Manometer with high accuracy and speed of response. The C9501/IS is ATEX Compliant and designed for use in areas where explosive or flammable substances are stored or processed.

  • Pressure Measurement Range: 0-to-±140 mBar.
  • Twin Inputs for Gauge or Differential Pressure.
  • Positive or negative (vacuum) pressure measurement.
  • Pressure Units: PSI, mbar, inH2O, mmHg, torr, Pa, mmH2O.
  • Adjustable zero value.
  • Maximum, minimum pressure and hold function.
  • Warranty: 2 Years

The Comark C9501/IS Intrinsically Safe Manometer has twin inputs for gauge or differential pressure measurement. With a durable design to survive an industrial setting, the Intrinsically Safe Pressure Meter 0-to-±140mbar is dust and waterproof to IP67 Standard (BS EN 60529), and an optional protective rubber boot is also available.

Our Intrinsically Safe Pressure Meters contain impregnated protection, designed to limit electrical discharge as well as the surface temperature of the device, in order to prevent the possible ignition of a volatile gases, dust or vapours.

HHSolutions_Comark_C9501_IS_Pressure_Meter

All Intrinsically Safe Models are supplied with MC95 Carry Case.

Owners of Intrinsically Safe Models should read this Product Information Bulletin

The Comark range of Intrinsically Safe Manometers are ATEX compliant, certified to Ex II 1G Ex ia IIC T4 Ga and Baseefa03ATEX0079.
The Comark C9500/IS Range are recommended for use in areas where explosive or flammable substances are stored or processed. These pressure meters have been approved by internationally recognised approval bodies to intrinsic safety requirements.

What is ATEX?
ATEX is the name commonly given to the two European Directives for controlling explosive atmospheres:

1) Directive 99/92/EC (also known as ‘ATEX 137’ or the ‘ATEX Workplace Directive’) on minimum requirements for improving the health and safety protection of workers potentially at risk from explosive atmospheres. The text of the Directive and the supporting EU produced guidelines are available on the EU-website. For more information on how the requirements of the Directive have been put into effect in Great Britain see the information in the section Explosive atmospheres in the workplace below.

2) Directive 94/9/EC (also known as ‘ATEX 95’ or ‘the ATEX Equipment Directive’) on the approximation of the laws of Members States concerning equipment and protective systems intended for use in potentially explosive atmospheres. The text of the Directive and EU produced supporting guidelines are available on the EU website. For more information on how the requirements of the Directive have been put into effect in Great Britain see the section on Equipment and protective systems intended for use in explosive atmospheres

 

Download Literature

    Comark Pressure Meters

    Operator Manual C9500 Series

Technical Specifications

NameValue
Pressure Measurement Range0 to ±140 mBar (See manual for other ranges)
Resolution0.1mBar
Accuracy±0.2% of full scale (FS) at +23°C
Repeatability±0.1% of full scale (FS) at +23°C
Mean Temperature Coefficient of Reading<0.1% per °C
Operating Range0 to 50°C, 10 to 90% RH non-condensing
Environmental ProtectionIP67, BS:EN60529, IEC529
Battery9V PP3 *6F22 — 6F22REL or GP1604G
125 hours
*IS models must only use Zinc Chloride Battery
Low Battery WarningDisplays BAT just before end point of battery
Over Pressure Indication– – – –
Over 9999 Reading IndicationLowest 4 digits blink
Connections1/8" BSP female
Weight330gm including battery
Warranty2 Years

Recently Viewed Products