InTemp CX602 Bluetooth 90 Day Single-Use Dry Ice Temperature Data Logger

Contact Us for Pricing:



The Onset InTemp CX602 Dry Ice data logger is a Bluetooth Low Energy (BLE)-enabled single-use logger that records temperature data for up to 90 days. The CX602’s built-in 1.5m probe measures temperatures as low as -95°C (-139°F) and includes a protective sheath to prevent accidental cutting of the cable.

Using the InTemp® app on your mobile device, you can easily configure the CX602 with one of multiple preset configurations, or set up a custom configuration. The logger has a built-in alarm to alert you to excursions. Logger reports can be quickly downloaded to your phone or tablet and uploaded to an InTempConnect® account, where you can analyse and export data as needed.

With the optional CX 5000 Gateway device, you can get real-time notifications via email or text of temperature excursions, and automatically download data from loggers within range of the device.

Supplied with a three-point ISO/IEC 17025 NIST 'UKAS' Certificate of Calibration, valid for one year from date of purchase.


KEY FEATURES OF THE CX602

  • Compact and lightweight – ideal for shipping pharmaceuticals
  • Bluetooth means no need to open containers to access data
  • Record up to 14,000 data points
  • Anti-cut cable sheath
  • International Air Transport Association (IATA) compliant battery

Integrates with mobile and cloud platforms for a 21CFR Part 11 compliant solution
Secure data via encrypted technology.

Download Literature

    Onset_InTemp_CX602_Data_Logger_Datasheet.pdf

    Onset_InTemp_CX600_and_CX700_Manual.pdf

    Onset_InTemp_CX600-CX700-QSG.pdf

Temperature Probe
Sensor Range -95° to 50°C (-139° to 122°F)
Accuracy ±1.5°C from -95° to -40°C (±2.7°F from -139° to -40°F)
±1.0°C from -40° to -10°C (±1.8°F from -40° to 14°F)
±0.5°C from -10° to 50°C (±0.9°F from -14° to 122°F)
Resolution 0.013°C at 25°C (0.023°F at 77°F)
Drift <0.1°C (0.18°F) per year
Response Time 1 minute 30 seconds typical to 90% in open air moving 1 m/s, unmounted
Calibration

Three-point NIST calibration, probe only

Cable Length 1.5 m (4.92 ft)
Probe Dimensions Food-grade stainless steel probe with blunt tip, 31.75 mm (1.25 inches) long, 4.78 mm (0.188 inches) diameter
Logger
Logger Operating Range -30° to 50°C (-22° to 122°F)
Radio Power 1 mW (0 dBm)
Transmission Range Approximately 30.5 m (100 ft) line-of-sight
Wireless Data Standard Bluetooth Low Energy (Bluetooth Smart)
Type Single use
Duration Up to 90 days
Shelf Life Up to 1 year
Logging Rate 1 second to 18 hours
Time Accuracy ±1 minute per month at 25°C (77°F)
Battery CR2450 3V lithium, not replaceable. Battery for the CX602 lasts for duration specified when logger is used within the 1-year shelf life. Faster logging intervals, remaining connected with the InTemp app, excessive downloads, and paging may impact battery life.
Memory 16,000 measurements 
Full Memory Download
Time
Approximately 30 seconds; may take longer the farther the device is
from the logger
Dimensions 6.9 x 4.5 x 1.1 cm (2.71 x 1.76 x 0.42 inches)
Weight 25.5 g (0.90 oz)
Environmental Rating IP65
CE Compliant The CE Marking identifies this product as complying with all relevant
directives in the European Union (EU).

Passed RTCA DO160G, part 21H

Dimensions 6.9 x 4.5 x 1.1 cm Batteries CR2450 Power Battery Only