InTemp CX402 CDC VFC Compliant Temperature Data Logger with Glycol Bottle
With the optional CX Gateway, you can get remote notifications of temperature excursions and automatically download data from loggers within range of the device. Learn more about the CX Gateway here.
Models (SELECT VARIANT) are available with a 2- or 4-meter external temperature probe attached to a Glycol bottle used to insulate the probe and delay thermal response times.
Please select preferred model from Variant Options above.
- Complies with all CDC and Vaccine for Children program requirements
- Easy-to-read display with current and daily min/max temperatures
- Bluetooth transmission of data and alarm states
- Real-time alerts and automatic data downloads with optional CX Gateway device
- Logger setup and download of data in seconds – without a PC or cable
- +/- 0.5C accuracy for reliable temperature readings
- Battery-powered to guard against local power loss
- Several Glycol bottle sizes with probe and cable-length options
- Includes 3 Year ISO 17025 compliant calibration
- Operating Temperature -30° to 70°C (-22° to 158°F)
- External Probe Range -40° to 100°C (-40° to 212°F)
Download Literature
HHSolutions-Onset-InTemp-VFC-data-sheetHHSolutions_InTemp_21_CFR_Part_11_Compatibility
HH_Solutions_InTemp-clinical-trials-data-sheet
Logger |
|
Radio Power |
1 mW (0 dBm) |
Transmission Range |
Approximately 30.5 m (100 ft) line-of-sight |
Wireless Data Standard |
Bluetooth Low Energy (Bluetooth Smart) |
Logger Operating Range |
-30° to 70°C (-22° to 158°F), 0 to 95% RH (non-condensing) |
Logging Rate |
1 second to 18 hours |
Battery Type |
Two AAA 1.5 V alkaline or lithium batteries, user replaceable |
Battery Life |
1 year, typical with logging interval of 1 minute. Faster logging intervals, remaining connected with the InTemp app, excessive report generation, numerous audible alarms, and paging all impact battery life. |
Memory |
128 KB (84,650 measurements, maximum) |
Full Memory Download Time |
Approximately 60 seconds; may take longer the farther the device is from the logger |
LCD |
LCD is visible from 0° to 50°C (32° to 122°F); the LCD may react slowly or go blank in temperatures outside this range |
Dimensions |
9.4 x 4.5 x 2.59 cm (3.7 x 1.77 x 1.02 inches) |
Weight |
90.2 g (3.18 oz) |
Environmental Rating |
IP54 |
The CE Marking identifies this product as complying with all relevant directives in the European Union (EU). |
External Temperature Probe |
|
Range |
-40° to 100°C (-40° to 212°F) |
Accuracy |
±1.0°C from -40° to -22°C (±1.8°F from -40° to -8°F) |
Resolution |
0.024°C at 25°C (0.04°F at 77°F) |
Cable Length |
2 or 4 meter (6.56 or 13.12 feet) flat ribbon cable |
NIST Calibration |
Single point NIST calibration, probe only |
Probe Dimensions |
Food-grade stainless steel probe with pointed tip, 53.34 mm (2.1 inches) long, 3.18 mm (0.125 inches) diameter |
Ambient Temperature Sensor |
|
Range |
-30° to 70°C (-22° to 158°F) |
Accuracy |
±1.0°C from -30° to -22°C (±1.8°F from -22° to -8°F) |
Resolution |
0.024°C at 25°C (0.04°F at 77°F) |